For
product enquiries please call +44 (0) 118
9311188
or email: info@able.co.uk |
|
The next generation of the renowned Jerome® family of vapour analyzers, the J605, maintains all the features, reliability and simplicity that have established Jerome® as an industry leader in the hydrogen sulphide and mercury vapour gas detection markets for more than 20 years.
Significant performance enhancements and new features now provide a new alternative for industrial hygienists and consultants to detect, measure and report levels of hydrogen sulphide vapour for environmental, industrial and waste water treatment measurements.
|
 |
|
The J605 has an accuracy of ±1 ppb at the calibration of 5 ppb - fullfilling the needs of the tightest of EPA requirements for hydrogen sulphide testing. The enhanced J605 has an incredible resolution of 20ppt and detects hydrogen sulphide vapour levels as low as 3 ppb with confidence. The J605 is housed in a lighter and more ergonomically designed case, setting a new precedence for low-level, portable monitors. The industry-proven, gold film sensor has been upgraded and improved, resulting in longer life expectancies. The larger display, onboard data logging, auto zeroing, and battery-powered sensor regeneration will make it easier to use and maintain the instrument in the field. Communication options and data acquisition have become more simplified with USB communication, SCADA compatibility via 4-20mA output, and menu-driven programming options.
| Jerome J605
Features |
- Rugged and easy to operate
- Battery powered sensor regeneration (requires external battery pack)
- Auto date/time stamp
- On board data logging 20,000 samples
- Inherently stable gold film sensor
- Pressure sensitive membrane switch operation
- Rechargeable internal battery pack for portability
- Wide detection range allows for multiple applications
- Automatic LCD backlight during low light conditions
- Survey mode for rapid source detection of hydrogen Sulphide concentrations
- Microprocessor ensures a linear response throughout the entire range of the sensor
- 4-20 mA output for SCADA systems
- No PC Software Required
- 20,000 Data Points
- Stores Date, Time and Locations
- On-Board Data Logging
- Optional USB interface for data transfer and printing capabilities
- 18 Hour Battery Life
- Detects H2S levels as low as 3 ppb
- Fulfills EPA requirements
|
| Optional
Accessories |
- Functional Test Module for instrument verification in the field
- Maintenance Kit for routine maintenance and upkeep
- Hard or Soft Field Carrying Cases for versatile handling and additional storage
- Car Adapter, External Battery Pack
|
| Applications |
- Leak Detection
- Quality Control
- Ambient Air Analysis
- Accuracy Check for other Hydrogen Sulphide Monitors and Control Systems
- Applied Research
- Scrubber Efficiency Testing
- Hydrogen Sulphide Source Detection
- Control Room Corrosion Monitoring
- Landfills
- Agricultural Applications
- Regulatory Compliance and Permitting
|
| Specifications |
| Resolution |
20 ppt in Range 0 |
| Detection
Range |
3ppb - 10 ppm |
| Accuracy / Precision |
Range 0: ± 1 ppb at 5 ppb / 10% RSD
Range 0: ± 3 ppb at 50 ppb / 5% RSD
Range 1: ± 0.03 ppm at 0.5 ppm / 5% RSD
Range 2: ± 0.3 ppm at 5.0 ppm / 5% RSD
|
| Response
Time |
Manual Range Selection
Range 2 (1.0 to 10.0 ppm) 12 seconds
Range 1 (0.10 to 1.0 ppm) 18 seconds
Range 0 (3 to 100 ppb) 12 to 27 seconds
Auto Range Mode 12 to 52 seconds
|
|
| Flow
Rate |
150 ml/min or 0.15
l/min |
| Power
Requirements |
12 VDC power adaptor |
| Internal
Battery Pack |
Rechargeable nickel metal hydride (NiMH) |
| Environmental
Range |
0-40 °C, noncondensing,
nonexplosive |
| Interface |
USB |
| Dimensions |
28cm L x 16cm W x 17cm H |
| Weight |
2.5 Kg |
| Warranty |
One year, factory
parts and labour |
| Certifications |
European Communities
(CE) and TUV: UL 61010 - 1:2004 standard and CSA - C22.2 NO. 61010 - 1 - 04 |
|
| |